Radcal ACCU-GOLD2 User Manual

Radcal ACCU-GOLD2 User Manual

For use with all accu-gold digitizers and accu-gold touch pro
Hide thumbs Also See for ACCU-GOLD2:
Table of Contents

Advertisement

Quick Links

dddddddd
ACCU-GOLD2
USER GUIDE
For use with
all Accu-Gold digitizers and
Accu-Gold Touch Pro

Advertisement

Table of Contents
loading

Summary of Contents for Radcal ACCU-GOLD2

  • Page 1 ACCU-GOLD2 USER GUIDE For use with all Accu-Gold digitizers and Accu-Gold Touch Pro...
  • Page 2 426 West Duarte Road Monrovia, CA 91016-4591 USA USA (626) 357-7921 Fax USA (626) 357-8863 email sales@radcal.com www.radcal.com Radcal Part # MNL/AG2 4094300 Rev: N Software V2.75 & ON Printed: Feb, 2022 Copyright © 2015, 2016, 2017, 2018, 2019, 2020, 2021, 2022...
  • Page 3: Table Of Contents

    Hardware Overview ............................. 7   Differences between the ‘legacy’ hardware and the ‘plus’ hardware ..............7   Digitizers ................................7   What is Accu-Gold2? ............................. 8   Starting Accu-Gold ..............................9   Starting Accu-Gold via Wi-Fi ..........................10   Nomenclature (Accu-Gold or Accu-Gold+) ....................... 11  ...
  • Page 4   Selecting a Display Profile ..........................26   Automatic Profile ..............................27   Custom Profile ..............................27   Data in List View..............................30   Exporting Data in List View ..........................30   Copying Data in List View ..........................32   Deleting Data in List View ..........................
  • Page 5   See additional discussion in Appendix C under Maximum Measurement Duration........53   Appendix A ................................54   Installing Accu-Gold ............................54   Installing the Calibration File ..........................55   Uninstalling Accu-Gold ............................56   Appendix B ................................57  ...
  • Page 6   Appendix E ................................94   Accu-Gold measurements and Feature comparison ..................94   Appendix F ................................97   Quiet-Bias Technology ............................97   Appendix G ................................98   Carrying cases – Accu-Gold ..........................98   Carrying cases – Accu-Gold Touch ......................... 100  ...
  • Page 7: Introduction

    Saving a Session For more in-depth information or troubleshooting, contact Customer Support at (626) 357-7921 extension 123 or send an email to Cust_Sup@Radcal.com see Appendix A for software installation.
  • Page 8: Hardware Overview

    Hardware Overview The Accu-Gold system consists of a digitizer module, sensor(s) and a display. The Accu-Gold digitizer is available in different models that are suited for specific applications, the differences being which sensors can be used. The Accu-Gold software is the same for all models but the measurement data that is available will depend on the model digitizer used and connected sensor(s).
  • Page 9: What Is Accu-Gold2

    What is Accu-Gold2? Radcal’s Accu-Gold software has been redesigned from the ground up with a modernized user interface and under-the-hood capabilities that will ensure continuing compatibility with Microsoft products (e.g. Windows & Office) as well as expanded capabilities for the Accu-Gold family of measurement systems.
  • Page 10: Starting Accu-Gold

    Starting Accu-Gold When launched, the Accu-Gold2 brings up a screen that lets you choose the appropriate connection method to the Accu-Gold digitizer. Connect via Wi-Fi – The Accu-Gold Nugget lets you connect to the digitizer wirelessly. Connect via USB - When the USB cable is the connection method to the digitizer.
  • Page 11: Starting Accu-Gold Via Wi-Fi

    For the Accu-Gold Touch, start the Touch in WiFi mode and wait for the blue light to flash rapidly. Start Accu-Gold2 and choose Connect via Wi-Fi: The software will look for “Nugget…” (or “Touch”) on the wireless networks list, and establish a connection to it sending it the password it requires.
  • Page 12: Nomenclature (Accu-Gold Or Accu-Gold+)

    Nomenclature (Accu-Gold or Accu-Gold+) AGDM – Accu-Gold Digitizer Module ADDM - Accu-Dose Digitizer Module RGDM – Rapid-Gold Digitizer Module AGMS – Accu-Gold Multi-sensor ICH – Ion Chamber DD - Dose Diode AGLS – Accu-Gold Light Sensor DAPCKP+ - DAP Check Plus The AGDM+ digitizer is compatible with the AGMS sensors (models AGMS-D and AGMS-M) but the AGMS+ sensors (models AGMS-D+, AGMS-M+, and...
  • Page 13: Main Screen Overview

    Main Screen Overview 1) Opens the Main Menu. 2) Lists all measurements made within the same session. You can scroll through each capture using the up and down arrows 3) Displays the connection method – USB or Wi-Fi Click the Connect [ ...
  • Page 14: Device Settings Menu

    13) Delete the active measurement. 14) Undo deleted measurement. 15) The Comments pane can be used to copy, cut and paste comments. Comments can be saved in the session along with any measurements and can optionally be carried over to the next measurement. The comments pane can be accessed from all views.
  • Page 15: Ic Wave Mode - On - Off

    system has not improved the experience, connect a Multi-sensor or Dose Diode and locate it somewhere in the radiation beam so that it can serve as a trigger source. High – Select if Std causes false triggering. Note: If you have a strong, noise free signal, selecting High will minimize the time the system measures a background zero in between measurements allowing you to make continuous measurements rapidly.
  • Page 16 Ion Chamber Exposure with incorrect stop trigger: Select Dynamic Stop Trigger Off: New exposure with Dynamic stop trigger Off:...
  • Page 17: Agms Pre-Pulse Mode

    The “AGMS Pre-Pulse Mode” algorithm is designed to accommodate the initial pre-heat kV pulses commonly found with some dental generators such as the Gendex 770. Pre-Pulse ON should not be selected for other pulsed waveforms such as pulsed fluoroscopy or tomography. Note: The dose remains integrated over the entire exposure regardless of the setting (On or Off), only the kV and time calculations are affected.
  • Page 18: Setting Up The Hardware

    Setting up the Hardware To set up the Accu-Gold hardware 1. Connect the sensor(s) to the digitizer. Connect the USB cable to the digitizer and then the computer. 2. Position the sensor to make a measurement. Note: Position the sensor before clicking the Connect button; moving the sensor or cable after clicking the Connect button may trigger a false measurement.
  • Page 19 The software begins to connect to the digitizer, as indicated by the status at the bottom of the screen: The software then begins to initialize the digitizer. If an ion chamber is connected, the bias supply needs to start and stabilize. 5.
  • Page 20 Note: If you attempt to quit Accu-Gold without saving or have made additional measurements since saving, Accu-Gold will prompt you with a “Save Session?” dialog box. Radcal strongly recommends that you always...
  • Page 21 Session files can be sent to Radcal as email attachments for rapid responses to any questions or difficulties you may have. You can continue to review the data in List or Wave view by clicking on the List or Wave buttons in the title bar.
  • Page 22: Comments

    Append snippet Minimize Copy Paste Carry-over pane can be used to enter notes or copy and paste details regarding the measurement. Comments are saved in the session along with any measurements and can optionally be carried over to the next measurement. The pane can be accessed from all views.
  • Page 23: Measurements

    Before making an exposure, make sure to choose the Anode/Filter combination that best represents the machine being measured. Radcal sensor calibrations used the following mammographic tubes and assume a 2.2 mm polycarbonate paddle, or simulated paddle (Appendix H), are placed on top of the sensor: Anode/Filter ...
  • Page 24: Making A Measurement With Multiple Sensors Connected

    W Ag  Selenia  Additional anode/filter combinations for specific machines are also included (see Appendix B) Making a Measurement with multiple sensors connected The Accu-Gold Digitizer Module (AGDM) allows you to connect up to five sensors and simultaneously collect data from all of the connected sensors. (The Rapid-Gold Digitizer Module (RGDM) and Accu-Dose Digitizer Module (ADDM) have a subset of these ports –...
  • Page 25: Long Exposures

    The maximum duration in this mode is 2.45 days. Note that sessions containing many long exposures with waveform data may take significant time to save or load data. Modern computers with fast disc storage and processors will have better performance in this situation. Contact Sales for Radcal’s current tablet display options.
  • Page 26: Measurement Limitations

    Measurement Limitations The Radcal AGMS Series of Multi-Sensors are highly capable instruments that will generally provide a comprehensive set of measurements throughout a wide range of diagnostic x-ray applications. However, there are some conditions under which the sensor will provide a more restricted set of measurements. These conditions...
  • Page 27: Selecting A Display Profile

    When making background or scatter measurements, the readings may be too low to trigger a measurement. In that case a second sensor placed in the beam can be used to trigger the measurement which will guarantee that the ion chamber’s signal is captured during the triggering period of that sensor. A 180cc, 500cc or 1800cc chamber may be used, depending on the radiation levels to be detected.
  • Page 28: Automatic Profile

    Automatic Profile In Measurement view, profiles define which values are displayed in the matrix. You will rarely want to choose a different profile other than Automatic Profile (Auto). Automatic Profile looks at the sensors in a measurement and whether it is diagnostic, fluoro or mammo and adjusts the matrix dynamically.
  • Page 29 Use the table preview popup to select the desired matrix layout. In the example above, the matrix changed from a 2x3 layout to a 3x3 layout. By changing to Custom Profile, the labels of the matrix fields become clickable. After clicking on the label, a drop-down menu appears for selecting a different result value.
  • Page 30 Each cell allows you to pick from the list of possible outputs from the digitizer and sensors. See Appendix E for a complete list. To create multiple custom profiles, click on the pencil in the lower right corner. The ‘Custom Profile’ becomes changeable.
  • Page 31: Data In List View

    Data in List View List view lets you see the measurement data in a table format. From this screen, you can perform the following actions: Exporting Data in List View Send all to Excel – opens a new instance of Excel and exports all of the measurements in the format that had been selected.
  • Page 32 Units – can be changed before exporting to Excel. Columns - Select the icon to expand the column chooser: The columns included in the export and the order they are displayed can be selected. When exporting the data, the units will occupy a separate column. If the Unit Position is set to Include in header, the data will be exported using the same number of columns as displayed.
  • Page 33: Copying Data In List View

    Copying Data in List View Selected cells can by copied to the clipboard via the button Copy or the hotkey Ctrl+C or via the context menu of the selected range. The header can be included or omitted by clicking on the arrow and selecting: Deleting Data in List View Individual measurements can be selected for deletion.
  • Page 35: Hide Individual Waveforms In Wave View

    “Low-pass Bandwidth Filter” to the waveform. The CT exposure you see below was made with a Radcal 0.15 cc thimble chamber placed in the central axial location of the Boone ICRU CT phantom. The scan started and ended with the beam beyond the end of the phantom.
  • Page 36   In the example below, the Lowpass Bandwidth Filter is OFF: Turning the filter on to 60 Hz dramatically reduces the noise and enhances the detail. It is now clear, exactly when the beam intercepts the phantom. The periodic beam attenuation by the patient couch is now  ...
  • Page 37: Region Of Interest" (Roi)

    Note: For short exposures (<1 s), filters should be applied with care so as not to smooth important detail. Because the filter is non-destructive, different values may be easily tried. Application of a filter will not affect any of the measurement values (e.g. dose, kV, time). When a session is saved with filters applied, Accu-Gold will retain the filtered waveform selection when the file is re-opened.
  • Page 38  Clicking on the magnifying glass in the ROI window will automatically expand the ROI to fill the Wave screen.  The copy results to clipboard and the Extract to new measurement from Region of Interest will be explained later in this document. As another example, assessment of performance should probably occur near the end of the exposure after the mA has stabilized.
  • Page 39: Copy "Region Of Interest" To Clipboard

    If the data contained in the ROI window does not provide all of the details needed for evaluation, clicking on the icon in the ROI window will put all of the details on the clipboard. A simple “Paste” command (CTRL-V), will put the text data in Excel or a text document. For the Example above: When a region of interest has been selected, it is possible to non-destructively extract an ROI selection and preserve it as a daughter of the original parent.
  • Page 40 Daughter designation Extracted Region    ...
  • Page 41: Agms Dm+, D+, M+ Will Display Dose Information When Filter Is Out Of Range

    >10% for filtrations exceeding 2mm Al. Note: there is no warning that the uncertainty has exceeded Radcal’s specification for dose once the 1 mm Al limit is reached. Examples with a W/Ag beam while adding aluminium. The ion chamber is a Radcal 6M:...
  • Page 42 Note: In the first exposure (above) Accu-Gold reported the hvl to be 600 um Al. In the exposure below, 600 um Al was added to the beam. The result was a reduction in dose of 0.50 mGy as reported by both the 6M ion chamber and the multi-sensor.
  • Page 44: Sessions

    Sessions Sessions allow you to perform multiple readings in one file. You can then review and save the information as necessary. Starting a Session To start a new session, perform the following actions: 1. On the Accu-Gold main screen, click the icon in the upper left corner and select New.
  • Page 45: Importing A Session

    2. Navigate to a saved session and click Open to open the session. Importing a Session Importing a session provides an easy way to combine measurement sessions. To import a session, perform the following actions: 1. On the Accu-Gold main screen, click and select Import Session.
  • Page 46: Exporting A Session

    2. In File name: type a name for the session. 3. Navigate to the desired folder and click Save to save the session. Exporting a Session To export a measurement session to a new Excel file, use the List View (page 30). To export a single measurement to Excel, select the measurement and click the Excel export button in Measure View or Wave View.
  • Page 48: Drag & Drop To Excel

    Drag & Drop to Excel In Measure view you may click on a data value and drag it directly to an open Excel sheet. You may also copy any individual Measure data item to the clipboard and then paste it in any open document.
  • Page 49: Excel Companion Mode

    Excel Companion Mode Excel Companion Mode helps you to keep Excel and Accu-Gold visible side by side. You can enter Excel Companion Mode from the Start Screen or from the Companion Mode icon in the title bar (near the minimize icon). On entering companion mode, Accu-Gold switches to the side of the screen and opens an instance of Excel next to it.
  • Page 50: Exporting To Excel In Companion Mode

    You can switch the format of the exported data by selecting it from the drop-down box:  Accu-Gold Format: The standard format with all necessary data produced by Accu-Gold. Radcal provides Excel templates for standard measurement tasks or sample templates that can be adjusted to your needs.  Nordic Format: This is the same export format as used by another vendor.
  • Page 51: Edit Columns

    When creating a template in Excel, we suggest that the AG2 data be automatically put into a separate sheet (e.g. DATA) and other cells would refer to it from there (see Video Tutorial available on Radcal.com\Support). If the data is being captured amidst other data on the sheet and it is...
  • Page 52: Nugget

    Nugget Using the Wi-Fi Nugget Using the Wi-Fi adapter to connect to the digitizer is very similar with a few exceptions. Turn on the Nugget before launching the program. When you choose to “Connect via Wi-Fi”, the program will look for Nuggets on the network.
  • Page 53    Select "Yes" for Is Wave Mode Enabled.  After recording a measurement, a preview of the waveform will be shown and will be automatically be  downloaded unless you turn off the ‘Auto Wave Download feature. If so, select ‘Download’ if required.   ...
  • Page 54: See Additional Discussion In Appendix C Under Maximum Measurement Duration

        The full waveform is now available from the "Measure" and "Wave" view.  Waveforms can be accessed for each exposure (any order) while the nugget is powered on. Once the Nugget or Touch is turned off, the waveform data will no longer be accessible. See additional discussion in Appendix C under Maximum Measurement Duration.
  • Page 55: Installing Accu-Gold

    Appendix A Installing Accu-Gold Installation notes:  Uninstalling old versions is not necessary.  This manual is on the flash drive in the Setup Content folder and is also available for viewing by selecting ‘Help’ from the menu.  Administrator privileges are required for installation. To install the Accu-Gold software, perform the following actions: 1.
  • Page 56: Installing The Calibration File

    5. After the installation is complete, click Close. Installing the Calibration File Calibration files for your multi-sensor will be imported during installation. If you receive a message that states that the calibration was not found, you can install the calibration manually. The legacy multi-sensor calibration data must be loaded from an external file and be reinstalled when it is recalibrated.
  • Page 57: Uninstalling Accu-Gold

    6. Click Open. The file will load and a message will appear that says “License file was imported.” Uninstalling Accu-Gold To uninstall Radcal Accu-Gold, perform the following actions: 1. Navigate to Add/Remove Programs in the Control Panel and select Accu-Gold 2.
  • Page 58: Sensor Specifications

    The Accu-Gold digitizer features multiple inputs for sensors. Notes: The time range is different when using a Nugget (or Touch). See Feature Comparison, Appendix D. Radcal mammographic sensor calibrations assume a 2.2 mm polycarbonate paddle, or simulated paddle (Appendix H), are placed on top of the sensor.
  • Page 60 Custom anode/filter calibrations for GE Range of Applications Anode/Filter Tube Filt (um) hvl (mm Al)* Paddle** Mo/Al GE DMR 22-49 1.0 mm 0.42-0.61 2.2 mm Rh/Al GE DMR 25-49 1.0 mm 0.42-0.61 2.2 mm Mo/Mo GE Pristina 22-32 16-44 0.24 -0.51 2.2 mm Rh/Ag GE Pristina...
  • Page 61 2-40 1.3-13.5 No Paddle Conditions: Tube position 0 degrees Bow-tie filter is OUT Generator in service mode Custom calibration for use with Radcal’s Model 8463D Pan-dental positioner Range of Applications Anode/Filter Tube Filt (mm) hvl (mm Al)* Paddle** W/Al Dental 60-100 1.5-4...
  • Page 62: Ich - Ion Chambers

    10X6-6 For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-6 is a general purpose in beam chamber with a well-documented wide dynamic range chamber with many dose and rate applications. It is also recommended for measuring exposure time in Auto Dose mode. The range of time (pulse width) extends from 10ms to 300s.
  • Page 63 10X6-6M For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-6M is a dedicated mammography chamber. The chamber is a world standard for mammography, ready for any mammographic tube track-filter combination. Extraordinary flat energy response over 10 - 40 keV has been documented in technical papers and makes corrections unnecessary.
  • Page 64 10X6-60 For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-60 is a service and image intensifier chamber. The dynamic range and thin profile is ideal for input dose at the image intensifier, high dose rates encountered in Fluoroscopy and Cine, spot film devices, and other special procedures. Chamber 10X6-60 Min Rate...
  • Page 65 10X6-60DAP For use with Accu-Gold the 10X6-60DAP is ideal for Dose Area Product (DAP) measurement of small beams such as Pan-Dental or CBCT-Dental. Chamber 10X6-60DAP Rate Specifications 11 pGy-m 110 Gy-m Exposure Specifications 56 pGy-m 290 mGy-m Exposure Rate Dependance <5% over rated rate specification Calibration Accuracy ±5% using X-rays @ IEC RQR-8...
  • Page 66 10X6-3CT For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-3CT is a Computed Tomography Dose Index (CTDI) and DWP or DLP Chamber. Although designed specifically for CT X-ray beam measurements, either free-in-air or mounted in a head or body phantom, it can be used for DWP and DLP* applications such as Dental X-ray measurements because of the chamber’s excellent energy and partial volume response as well as uniformity along its entire 10 cm active length.
  • Page 67 10X6-10 This 10cc multi-purpose ion chamber is suitable for measuring R&F as well fluoro. It’s small size (5cm diameter x 1cm thick), excellent energy response and x-ray transparency makes it ideal for those difficult measurements where space is at a minimum. Chamber 10X6-10 Min Rate...
  • Page 68 10X6-180 For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-180 is a leakage and low level measurements chamber. The chamber is designed for leakage measurements with a cross-section of 100 cm (complies with CFR21) and volume of 180 cm . It is also used for very low dose to image receptors. Chamber 10X6-180 Min Rate...
  • Page 69 10X6-1800 For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-1800 is a radiation protection chamber used for low-level radiation measurements such as shielding leakage, irradiator, and environmental. Unlike typical survey meters, the 1800cm volume chamber offers improved accuracy over a wider dynamic range.
  • Page 70 10X6-500 Leakage Measurement Chamber A single sensor leakage measurement solution for very low-level radiation measurements such as shielding, leakage, irradiators and environmental that is part of the Accu-Gold family of systems. Equivalent to the Fluke® 96010A Ion Chamber used in the Fluke® Radiation Leakage Detection System, this chamber meets 21 CFR 1020.30(k) for leakage measurement requirements.
  • Page 71 10X6-0.6 For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-0.6 is a high dose rate chamber. This high dose rate chamber provides an excellent response at therapy and other high energy, high dose rate applications. The fully guarded chamber is mounted at the end of a 12-m low noise triax cable. It can be used with or without a build-up cap depending upon the application.
  • Page 72 10X6-0.6CT The 0.6cc thimble ionization chamber is ideal for dose measurements in modern wide beam multi-slice CT. It is calibrated using X-rays @ 150 kVp. Phantom adapter is included. Chamber 10X6-0.6CT Min Rate 20 R/s 200 nGy/s Max Rate 133 R/s 1.17 Gy/s Min Dose 100 R...
  • Page 73 10X6-0.18 For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-0.18 is a high dose rate chamber used for in-beam measurements of high-intensity gamma radiation. The fully guarded chamber is mounted at the end of a 3-m low-noise triax cable. Chamber 10X6-0.18 Min Rate 50 R/s...
  • Page 74: Ma/Mas

    90M9-AG Invasive Sensor The 90M9-AG is an invasive sensor. The mAs sensor connects to the 4-pin connector labeled mAs on the digitizer using a 4-m cable and to the X-ray generator with safety shrouded banana jacks. It is intended to be inserted in series with the X-ray generator return.
  • Page 75 Extraneous signals are generally not a problem for tube currents above about 100mA. For measurements where errors in the 5 mA range are significant, Radcal recommends the following:...
  • Page 76 1. Position the sensor on the anode cable so the cable doesn’t move relative to the sensor. 2. Locate the sensor away from magnetic noise sources, particularly those that might change during the exposure because the zero correction is measured several seconds before the exposure.
  • Page 77: Dd - Solid State Dose Sensors

    DDX6-W (DISCONTINUED) The DDX6-W is a solid-state dose sensor for diagnostic range measurements. It is a small solid-state sensor for single exposures and fluoroscopy in the diagnostic range. It measures dose, dose-rate, and time and is passively corrected for beam quality. It may be used simultaneously with an AGMS-D multi- sensor in entrance / exit exposure measurements.
  • Page 78 DDX6-WL For use with Accu-Gold, Accu-Pro, and Accu-Dose, the DDX6-WL is a solid-state dose sensor for diagnostic range measurements. It is a small solid-state sensor for single exposures and fluoroscopy in the Diagnostic range. It measures dose, dose-rate, and time and is passively corrected for beam quality. It may be used simultaneously with an AGMS-D multi-sensor in entrance / exit exposure measurements.
  • Page 79: Aux Sensors

    AGLS Light Sensor The AGLS light sensor contains a silicone photo diode that can be exposed to ambient light through an optical filter, which very accurately simulates the spectral response of the human eye’s photopic response, in order to measure illuminance in lux units. A luminance adapter can be screwed onto the base unit in front of the photo-diode for the measurement of luminance in cd/m2 units through an aperture with a 10- mm diameter.
  • Page 80: Legacy Sensors

    Legacy Sensors The Accu-Gold+ family of products offer a number of enhancements over the Accu-Gold products they replace. These include:   Support for the AGDM+ multi-sensors providing simultaneous support for diagnostic and mammographic measurements.   Improved sensitivity at lower dose rates consistent with the industry trend as imaging systems require less radiation to form excellent imagery.
  • Page 81: Agms - Solid State Multi Sensors

    AGMS – M (DISCONTINUED) For use with Accu-Gold, the AGMS-M is a Solid-State Dose multisensor for Mammographic Range Measurements. It measures dose, dose-rate, time, kVp with FFT analysis, Flash HVL, and beam filtration. It features Flash correction of dose for beam quality. Type Anode / Filter Tube...
  • Page 82 For use with Accu-Gold, the AGMS-D is a Solid-State Dose multi-sensor for Diagnostic Range Measurements. It is used for single exposure and fluoroscopy. It measures dose, dose-rate, time, kVp with FFT analysis, Flash HVL, and beam filtration. It features Flash correction of dose for beam quality. Type hvl (mm AI) ±10%...
  • Page 83: Nugget - Accu-Gold Wireless Adapter

    Accu-Gold software. It uses a built-in Lithium-ion polymer battery with battery management circuitry that maintains the life of the battery by carefully controlling the charging while monitoring its temperature and voltage. Radcal Part Number Accu-Gold Nugget MDL/AGNUGGET Charger (5V, 2.0A)
  • Page 84: Led Blinking

    Power Switch Status Meaning Solid Green Power On Blinking Green Low Battery (~25% remaining) Blinking Yellow Battery Charging Solid Yellow Battery Fully Charged Long on-off pause (4 sec) Battery Error Wi-Fi Status Meaning Orange Starting Blinking Green Wi-Fi Ready Error Additional indications: Fast blinking orange is used to indicate firmware update has been completed.
  • Page 85: Operation

    For information on installing the software, see Appendix A. Snap the Nugget onto the digitizer module: Align the USB connectors and slide the Nugget all the way in. (To remove it, pull out the connector first then unsnap it by tilting it upward or sliding it off the end of the digitizer.) Plug in a multi-sensor, ion chamber or...
  • Page 86 Connecting manually If the Nugget is not able to connect automatically, you will need to connect manually. If there are multiple Nuggets that show up, you will also have to connect manually. For Windows 10, if you get this dialog box, click on “Connect using a security key instead”.
  • Page 87 NOTE: Internet Protocol Version 4 (IPv4) must not be set to a fixed IP address. IPv4 must be set to “Obtain address automatically” and “Obtain server address automatically. Contact us if any issues arise.
  • Page 88: Battery

    3.7 V Li-ion polymer 1150mAh single cell - not removable. Battery life will depend on how it is being used. Turn off the unit when it is not being used to conserve battery life. Turning the unit back on will require the unit to re-establish the Wi-Fi connection which may take a minute or so at most.
  • Page 89: Physical Description

    Dimensions 135 mm x 62 mm x 33 mm (5.3" x 2.4" x 1.3") Weight 0.12 kg (0.25 lbs) 15 °C to 35 °C Operating temperature: Pressure: 60 to105 kPa Humidity: Up to 80% RH or 20 g/m -20 °C to +60 °C Storage Temperature: Network Standard Support: IEEE 802.11b/g/n Frequency band: 2.400 - 2.472 Ghz, channels 1-11...
  • Page 90: Nugget Measurement Feature Comparison

    Using Accu-Gold V2 software Feature Nugget Support for AGDM Support for AGDM+ 10 kHz Signal Sampling Support for AGMS-D Sensor Support for AGMS-M Sensor Support for AGMS-DM+ Sensor Support for AGMS-D+ Sensor Support for AGMS-M+ Sensor Support for Ion Chambers Support for Dose Diodes Support for Light Sensor Support for Current Clamp...
  • Page 91: Firmware Update

    Duty Cycle Results Pulse Frequency Ave Pulse Width Result Max Pulse Width Result not planned Min Pulse Width Result not planned Dose / Pulse Result Pulse Count IC Temperature Result IC Air Pressure Result Reads temp and pressure on start Updates calibrations, temp and future pressure for each measurement...
  • Page 92: Regulatory

    Press the power-on button of the Nugget. The Wi-Fi light will come on GRN then ORN. When it starts blinking orange quickly, the update is complete. Unplug the flash drive, press the Nugget power button to turn it off and then back on again. It is now ready to use. EU (ETSI): This device is intended for home and office use in all EU countries (and other countries following the EU directive 1999/5/EC).
  • Page 93 installed to provide a separation distance of at least 20 cm from all persons and must not be co- located or operating in conjunction with any other antenna or transmitter. Industry Canada Statement: Contains: IC: 8407A-RSWC201 This device complies with RSS-210 of the Industry Canada Rules. Japan (Telec) Contains: Telec Construction Certification Number 001-A13582.
  • Page 94: Accu-Gold Touch

    Appendix D Accu-Gold Touch The Accu-Gold Touch USB mode and WiFi mode mimic the computer connected digitizer and WiFi connected Nugget operation. The regulatory specifications of the Nugget apply to the Touch. Measurements exported from the Accu-Gold Touch can be read by Accu_Gold 2 Version 2.46 and on. All functions for displaying and manipulating measurement session data are available for the Touch measurement sessions.
  • Page 95: Accu-Gold Measurements And Feature Comparison

    Appendix E Accu-Gold measurements and Feature comparison The following are measurements that can be assigned to any cell when creating custom profiles. These are grouped by the measurement source. Note that not all measurements are available when using the Nugget or Touch WiFi mode. USB Cable/ Nugget/ Definition...
  • Page 96 Dose / Total Dose Ratio of the measurements of extracted dose to total dose. Dose Rate Dose Rate Dose Ratio AGMS / DD Multi-sensor dose divided by dose-diode dose. Dose Ratio AGMS / IC Multi-sensor dose divided by ion-chamber dose END is useful for capturing the radiation values at the end End Average kV of an exposure.
  • Page 97 Dose / Total Dose Ratio of the measurements of extracted dose to total dose. Dose Rate Dose Rate Dose Ratio DD / AGMS Dose Ratio DD / AGMS Dose Ratio DD / IC Dose Ratio DD / IC Total Dose Total dose of source measurement.
  • Page 98: Quiet-Bias Technology

    Special grounding accessories partially mitigated this effect, but were often inconvenient. Radcal is excited to announce that research and development efforts have yielded a new ion chamber technology, the “Quiet Bias”. A specially designed electronic circuit filters the bias voltage supply eliminating noise that would have appeared in the output signal.
  • Page 99: Carrying Cases - Accu-Gold

    Appendix G Carrying cases – Accu-Gold Soft Cases – Model AGSC...
  • Page 100 Hard Cases: 40V94S Single Layer Carry Case 4” (100 mm) deep 40V94R Accu-Gold Carry Case 2-layer, 5” (125 mm) deep...
  • Page 101: Carrying Cases - Accu-Gold Touch

    40V94TP Accu-Gold Carry Case 3-layer 150 mm deep Carrying cases – Accu-Gold Touch 40V94E Accu-Gold Touch Soft Carry Case 1-layer 125 mm deep...
  • Page 102 40V94D Accu-Gold Touch Hard Carry Case 1-layer 125 mm deep 40V94DP Accu-Gold Touch Water-proof Carry Case 2-layer 150 mm deep...
  • Page 103: Accessories

    Appendix H Accessories ---- Note for Mammography Sensor users. The Mammographic Sensors, models RAPD-M, AGMS-M, AGMS-DM+ and AGMS-M+ are calibrated using a 2.2 mm thick polycarbonate compression paddle placed in beam for all measurements (kVp, dose, hvl). If you are making measurements where it is not practical to use the machine’s paddle in beam, use this 2.2 mm polycarbonate plate (Model 8154 - Simulated Paddle) on top of the sensor during your measurements.
  • Page 104: 8463D - Multi-Sensor Positioner For Pan-Dental Machines

    The Model 8463D Pan-Dental Positioner has been designed to facilitate attaching and reliably positioning an AGMS-D+ or –DM+ multisensor on many types of panoramic dental systems. Using the Accu-Gold 2 software, a correct dose measurement can be obtained...
  • Page 105 Test Stands and Positioners – ADTS-W – Accu-Dose Test Stand AGTS-W – Accu-Gold Test Stand...
  • Page 106: Appendix I - Warranties And Disclosures

    Radcal, with shipping costs prepaid by the customer. Costs to return-ship to customer by ground transportation will be paid by Radcal if the repairs are warranty-applicable. This warranty excludes batteries.
  • Page 107: Data Loss

    WEEE and RoHS Accu-Gold meets the requirements of the . Radcal has implemented full compliance. Recycling manuals are available on request. Accu-Gold/Rapid-Gold/Accu-Dose+ meets the requirements of the EU-RoHS directive for RoHS 3, The Restriction of the Use of Certain Hazardous Substances (RoHS) in Electrical and Electronic Equipment Directive (EU Directive 2015/863) .
  • Page 108: Eu Declaration Of Conformity

    EU Declaration of Conformity Manufacturer’s Name: Radcal Corporation Manufacturer’s Address: 426 West Duarte Road Monrovia, CA 91016 U.S.A. Product Name: Accu-Gold, Rapid-Gold, Accu-Dyn Model Number(s): AGDM+, RGDM+MA, ADDM+, AGDN+ Product Accessories: Is in conformity with the provisions of the following CE directives: ...
  • Page 109 UK Declaration of Conformity Manufacturer’s Name: Radcal Corporation Manufacturer’s Address: 426 West Duarte Road Monrovia, CA 91016 U.S.A. Product Name: Accu-Gold, Rapid-Gold, Accu-Dyn Model Number(s): AGDM+, RGDM+MA, ADDM+, AGDN+ Product Accessories: AGMS-x, 10X6-x, 90Mx, DDX6-WL Is in conformity with the provisions of the following UK directives: ...