GE Healthcare TONOPORT V Ambulatory Blood Pressure System Firmware Version FW 2.1 Hardware Version HW 2.1 Service Manual 2001589-094 Revision D...
Initial release 2012-01 Chapter 4: TONOPORT V, V2.0 corrected to TONOPORT V, V2.1 Chapter 5: 2 pictures for TONOPORT V, version HW2.1 Rev. B added and difference between Rev. A and Rev. B explained Chapter 6: Required Tools: Windows 7 added Chapter 8.2: 1 picture for TONOPORT V, V2.1 Rev.
Introduction: Safety Information Safety Information 1.2.1 Responsibility GE Healthcare is responsible for the effects of safety, reliability, and performance only if: Assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by GE Healthcare. The electrical installation of the relevant room complies with the requirements ...
Any unauthorized attempt to repair equipment under warranty voids that warranty. It is the user’s responsibility to report the need for service to GE Healthcare or to one of their authorized agents. Failure on the part of the responsible individual, hospital, or institution using ...
Page 8
Introduction: Service Information For your notes TONOPORT V Revision D 2001589-094...
OFF Info Start mmHg unit is mmHg Info unit is kPa Measurement limits - Max. cuff inflation time = 130 seconds - Max. measurement time after start of measurement = 180 seconds Revision D TONOPORT V 3-1 2001589-094...
Page 12
For your notes TONOPORT V Revision D 2001589-094...
9600 Baud / 8 Bit / 1 Stopbit none Parity Cable USB TONOPORT V - PC (2001589-040) - not shown This cable connects the USB port of TONOPORT V, V2.1 (USB B-mini, male, 5-pin) to the USB port (USB A, male) of the PC. Revision D TONOPORT V 4-1...
Page 14
For your notes TONOPORT V Revision D 2001589-094...
Pictures Pictures of TONOPORT V for version HW2.1, Rev. A (GE part number: 2001589-092, Rev. A) Revision D TONOPORT V 5-1 2001589-094...
Page 16
(GE part number: 2001589-092 Rev. B) NOTE TONOPORT V, HW2.1, Rev. B uses a main board with another type of deflation valve than Rev. A. Difference between Rev. A and Rev. B: The main boards of Rev. A and Rev. B differ in the deflation valves used. The defla- tion valves are tightly connected with the main boards in Rev.
TONOPORT V version and loads the corresponding firmware for TONOPORT V, HW2.1: firmware FW2.1 or higher version. NOTE If the update program detects an older TONOPORT V version, it loads for TONOPORT V, HW2.0: firmware FW2.0 and for TONOPORT V, HW1.4 and earlier: firmware version FW1.4...
Firmware Loading Procedure: Loading Procedure Loading Procedure Connect inactive TONOPORT V to USB port or RS232 port via the USB or RS232 cable. Turn on TONOPORT V. NOTE After TONOPORT V has been switched on, updating can only be carried out within the first 10 minutes.
Page 19
Firmware Loading Procedure: Loading Procedure Check the correct firmware version on TONOPORT V: Step Action Diplay Switch on TONOPORT HH:MM (time) Press “INFO” 5 times Press “START” 21 : 00 (21 = Version FW 2.1 00 = no direct printout) Press "START"...
Page 20
Firmware Loading Procedure: Loading Procedure For your notes TONOPORT V Revision D 2001589-094...
E 14 Diastolic reading below 40 mmHg. Code appears when the cuff pressure has dropped to 40 mmHg and no diastolic pressure could be identified (TONOPORT V does not measure diastolic pressures below 40 mmHg). E 15 Motion artifact during diastole detection.
Page 22
At the same time the device switches to a deflation rate of 4 mmHg/s. When it detects more than 13 oscillations later on, the rate is changed to 6 mmHg/s. TONOPORT V Revision D 2001589-094...
Fixed volume of 0.5 L as an air container T-adapter Step-by-Step Instructions Only the pressure amplifier needs to be calibrated. This is done after switching on the TONOPORT V and waiting until the time is displayed on the LCD. Revision D TONOPORT V 8-1 2001589-094...
Page 24
Use a screw driver to turn the trimming potentiometer R53 until the LCD indicates a value of 50 ±20. Connect a reference pressure gauge, a rubber bulb and air container (0.5 L) to the TONOPORT pressure connector and press "START". TONOPORT V Revision D 2001589-094...
Page 25
Using a screwdriver, turn the trimming potentiometer R40 until the LCD indicates a value of 100. Also check the values indicated on the LCD at the following pressures (max. permitted deviation for each measurement: ±2 mm Hg): 50 mmHg Revision D TONOPORT V 8-3 2001589-094...
Page 26
Adjustment Instructions: Step-by-Step Instructions 150 mmHg 200 mmHg Switch the TONOPORT off. TONOPORT V Revision D 2001589-094...
Adjustment Instructions: Main Board - View of the Components Main Board - View of the Components Main Board, Rev. A (GE part number 2002866-005, Rev. A) for TONOPORT V, HW2.1, Rev. A TP9 TP10 Revision D TONOPORT V 8-5...
Page 28
Adjustment Instructions: Main Board - View of the Components Main Board, Rev. B (GE part number 2002866-005, Rev. B) for TONOPORT V, HW2.1, Rev. B TP9 TP10 TONOPORT V Revision D 2001589-094...
Transport and storage Temperature -20...+ 70 °C Relative humidity 10...90 %, no condensation Atmospheric pressure 500 hPa (altitude +4500 m relative to sea level) to 1060 hPa (-400 m relative to sea level) Revision D TONOPORT V 9-1 2001589-094...
Page 30
Technical Specifications: Dimensions Height 27 mm Width 80 mm Depth 100 mm Weight 199 g with batteries TONOPORT V Revision D 2001589-094...
Measure the time between start running the pump and stopping the pump. The time must be 130 sec max. If the time is over this value, the TONOPORT V must be sent in for inspection. Revision D TONOPORT V 10-1...
Press START: display indicates internal value, which must be 50 ±20. If the value lies outside this range, the TONOPORT V must be sent in for inspection. Press START again: display indicates "0" (the device is now in the calibration ...
Checkout Procedure After repair or maintenance of the TONOPORT V, a checkout procedure is recom- mended. This checkout procedure consists of the adjustment of the device, see Chapter 8, “Adjustment Instructions” the technical inspection of the measuring system, see Chapter 10, “Technical ...
Page 34
Overpressure test passed? o Yes o No Time limit test passed? o Yes o No Visual inspection passed? o Yes o No Checkout OK? o Yes o No Date:......Name and signature of test person: ................11-2 TONOPORT V Revision D 2001589-094...
TONOPORT V Operator Manual Serbian 2001589-111 TONOPORT V Operator Manual Estonian 2001589-112 TONOPORT V Operator Manual Rumanian 2001589-113 TONOPORT V Operator Manual English US 2001589-127 TONOPORT V Operator Manual Indonesian 2001589-128 TONOPORT V Operator Manual Croatian Revision D TONOPORT V 12-1...
TONOPORT V Operator Manual Latvian 2001589-131 TONOPORT V Operator Manual Lithuanian 12.2 Accessories Part No. Description 2001589-011 Cable RS232 TONOPORT V ———> PC (with CardioSoft V4.14 or higher) 2001589-040 Cable USB TONOPORT V ———> PC (approx. 1.5 m) 2001589-015 Carrying Bag 2001589-016...
PCB HW2.1 (with Motorola Pressure Sensor and LCD) (see pictures in chapter 5 and chapter 8.2 of this Service Manual) 2032506-001 Speaker TONOPORT V 12.4 Exchange Units Part No. Description 2003386-003 Exchange Recorder ABP TONOPORT V, V2.1 Revision D TONOPORT V 12-3 2001589-094...
Page 38
Parts Lists: Exchange Units For your notes 12-4 TONOPORT V Revision D 2001589-094...
Page 40
Fax: +49 30 2138542 Asia Headquarters GE Medical Systems GE Medical Systems Information Technologies, GmbH Information Technologies, Inc. GE (China) Co., Ltd. No1 Huatuo Road, Munzinger Straße 5 8200 West Tower Avenue Zhangjiang Hi-Tech Park Pudong, Milwaukee, WI 53223 USA 79111 Freiburg Shanghai, P.R.China 201203...